      Location:HOME INDUSTRY I... High-Frequency Electrosurgical Unit (ESU) Operating Procedures, Key Points, and Precautions

      High-Frequency Electrosurgical Unit (ESU) Operating Procedures, Key Points, and Precautions

      2025-05-12

      High-frequency electrosurgical units, as indispensable medical devices in modern surgical procedures, utilize high-frequency electrical currents to achieve tissue cutting and coagulation functions, significantly enhancing surgical efficiency while reducing intraoperative bleeding. However, improper operation may pose safety risks. Simon?Medical will provide standardized guidelines for medical staff from the following aspects: usage protocols, key operational considerations, and medical personnel requirements.

      ?

      I. Usage Protocols
      Preoperative Preparation

      Device Inspection: Ensure proper connections of the main unit, foot pedal, and electrode cables; verify the power supply is functioning properly; inspect disposable electrode pads for integrity.

      Environmental Safety: Remove flammable materials (e.g., alcohol-soaked cotton balls) near the surgical bed; ensure the operating table is dry and the patient is insulated from metallic surfaces.
      配件1.jpg

      Electrode Pad Placement

      Site Selection:

      Apply electrode pads to muscle-dense areas near the surgical site (e.g., thigh, buttock) after shaving the area; ensure firm adhesion and a contact area?≥100 cm2.

      Precautions:

      Avoid placement near bony prominences or scar tissue to prevent current concentration and thermal injury.

      Parameter Settings

      Mode Selection:

      Choose operating modes (Cutting/Coagulation) based on surgical type and tissue characteristics.

      Initial Power Recommendations:

      Cutting Mode: 40-60W (Note: Power output of Simon?Medical high-frequency electrosurgical units is adjustable in real time).

      Coagulation Mode: 60-80W, adhering to the?"low-to-high" principle?for incremental adjustment.

      Special Considerations:

      Reduce power output for pediatric or debilitated patients to avoid excessive tissue damage.

      Operational Testing

      Pre-activation Check:

      Perform a brief activation using?moist gauze?or a?dedicated test probe?to verify device functionality.

      Verification Criteria:

      Observe?spark intensity?and?auditory feedback?for normal operation;

      Confirm?absence of alarm signals?or system abnormalities.
      2.jpg

      Postoperative Procedures

      Device Shutdown:

      Power off the device, remove electrode pads, and inspect the patient's skin condition?for signs of electrical burns or complications.

      Equipment Maintenance:

      Clean the cutting tip thoroughly, sterilize it, and store appropriately for future use;

      Return the main unit to its designated storage location and?document operational parameters?for traceability.

      ?

      II. Key Operational Guidelines and Risk Mitigation
      Safety First

      Fire Hazard Prevention:

      Strictly prohibit use?in environments with flammable substances (e.g., alcohol). Electrosurgical sparks may ignite explosive vapors.

      Patient Safety Protocols:

      Ensure?complete insulation?between the patient and metallic surfaces to prevent?current diversion burns;

      Never reuse disposable return electrodes?or reposition detached pads during procedures.

      ?

      Power Adjustment Principles

      Gradual Increment Protocol:

      Always initiate at the lowest effective power?and incrementally increase following the?"low-to-high" principle?to minimize thermal tissue damage.

      Special Tissue Considerations:

      Increase power appropriately?when operating in deep tissues or fluid-rich environments (e.g., abdominal cavity) to maintain efficacy.

      Intraoperative Monitoring

      Device Alert Response:

      Immediately suspend operation?upon triggering of alarms (e.g., poor contact, power overload) and?conduct systematic troubleshooting.

      Patient Response Assessment:

      Cease activation immediately?if abnormal muscle fasciculations or cutaneous erythema occur; verify circuit integrity and electrode contact.

      ?

      Special Population Considerations

      1. Absolute Contraindications

      Cardiac Pacemaker Patients:

      High-frequency electrosurgery is strictly contraindicated?in patients with implanted cardiac pacemakers. Electrosurgical currents may:
      ○ Disrupt pacemaker function;
      ○ Induce myocardial injury via pacemaker lead thermal transfer;
      ○ Potentially trigger?ventricular fibrillation.

      2. Metallic Implant Precautions

      For patients with retained/prosthetic metallic objects (e.g., shrapnel, orthopedic hardware, joint replacements, stents):

      Electrode Placement Strategy:
      ○ Position return electrodes to establish current pathways?away from metallic implants;

      Operational Modifications:
      ○ Utilize?lower power settings;
      ○ Limit activation duration;
      ○ Enhance heat dissipation protocols to prevent?eddy current-induced thermal injury.

      3. Patient Positioning Requirements

      Immobilize extremities?to prevent contact between body parts with differential electrical potentials, particularly?small-area contact points?which may concentrate current density and cause burns.

      4. Pediatric Population

      Mandatory Use?of?pediatric-specific electrode pads?with optimized contact geometry during infant/neonatal procedures.

      III. Healthcare Personnel Requirements

      1. Mandatory Training Standards

      Qualification Prerequisite:

      Operators?must undergo comprehensive training?demonstrating proficiency in:
      ○?Operational principles?of high-frequency electrosurgical units;
      ○?Standardized protocols?for device application and safety management.

      2. New Device Implementation Protocol

      Prior to using?newly introduced electrosurgical models:

      Mandatory Documentation Review:
      ○ Thoroughly study?operation manuals?and?safety advisories?issued by Simon?Medical;

      Preclinical Validation:
      ○ Conduct?full-functional testing?in non-clinical settings to verify performance parameters and alarm responsiveness.

      ?

      Personal Protective Equipment (PPE) Requirements

      Insulated Gloves:

      Wear?high-integrity rubber gloves?with certified dielectric properties during all procedures.

      Footwear Specifications:

      Use?dry, thick-soled footwear?demonstrating verified electrical resistance to maintain grounding isolation.

      Operational Safety Objective:

      Ensure?continuous electrical isolation?from both grounded surfaces and the electrosurgical unit to establish?effective self-protection.

      ?使用指導(dǎo).jpg

      ?

      Summary
      The clinical efficacy of high-frequency electrosurgical units (HF-ESU) is contingent upon?strict adherence to standardized protocols?and?meticulous execution of operational details. Medical personnel must:

      Maintain?strict compliance with established procedures;

      Exercise?enhanced intraoperative vigilance;

      Implement?context-driven parameter optimization?to achieve?optimal surgical outcomes?while?maximizing patient safety.

      Simon?Medical strongly recommends:

      Periodic preventive maintenance programs?to ensure device reliability;

      Mandatory operator competency training?to mitigate clinical risks and?maximize the therapeutic value?of electrosurgical systems.

      ?

      Contact for technical support: 138-2067-8142

      ?


      previous page: AI大模型翻譯 Focus on the forefront of healthcare, and join us for the golden autumn event - the 92nd China International Medical Equipment Fair (CMEF Autumn) is about to kick off!

      Next page: What types of surgeries is high-frequency electrosurgical knife suitable for?

      PHONE:
      13820678142
      微波治療儀_高頻電刀-賽盟醫(yī)療科技
      Tianjin SIMON medical technology Co.,Ltd

      Tel: 86 22-83712046/?83712048

      Mobile: +86 166 2215 4239

      ???????????? +86 138 2067 8142

      Email: tjsaimeng@126.com

      Website: http://www.yujt.cn

      Address: Room 305-1, Building D, No. 6 Zhuyuan Road, Huayuan Industrial Zone, Tianjin

      日韩成人AV影视|免费观看av三级|91av三级在线|一级黄色影片|成人一级黄色片无码|欧美色图图片婷婷五月天|成人69
      <code id="ongbw"></code>

        1. <option id="ongbw"></option>